Abstract: AS 2879.9:2002 pdf download.Alumina Part 9: Determination of flow time. 6 SAMPLE PREPARATION The analytical sample shall be equilibrated to the laboratory atmosphere for a minimum of 2 h in a layer of 5 mm maximum thickness. The sample shal...
AS 2879.9:2002 pdf download.Alumina Part 9: Determination of flow time.
6 SAMPLE PREPARATION
The analytical sample shall be equilibrated to the laboratory atmosphere for a minimum of 2 h in a layer of 5 mm maximum thickness. The sample shall then he split as per AS 4538.2 into three portions of approximately 105—120 g for testing. The sample shall then be screened through a sieve (5.5) and the oversize material discarded.
7 PROCEDURE
7.1 Gener1
This procedure is for manually timed determinations.
7.2 Number of determinations
Three determinations shall be carried out.
7.3 Determination
Each determination shall be carried out as follows:
(a) Set up the apparatus in a vibration-free environment.
(b) Ensure that the funnel is secure on a stand and that the upper surface is precisely level.
(c) For apparatus not in frequent use. pre-condition the funnel by passing several portions of alumina through it. Discard this
material prior to testing.
(d) Accurately weigh 100 ±0.1 g of the prepared sample into a suitable container. (See Clause 6.)
Recommended: